Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 118

Advertising
background image

IM 12B07D02-01E

12-6 Appendix

2/3

QIS 12B07D02-21E

3.3 Current Output Test

Connect the instruments as shown in Figure 1, and set them as follows.

Connect a shorting bar between terminals LOW and IMP.

Decade resistance box:

921.6 :

Standard voltage/current source:

0 mV

DC source:

24 V DC

Use the following procedure to enter the inspection mode.
(1) Entering Service Code 87

a. Press the [ ] key. The message display will show “ OUTP.”

b. Press the [NO] key until the message display shows “ SERV.”

c. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing.

d. Press the [š] key eight times. The data display will show “80” with the first digit of 8 flashing.

e. Press the [>] key once. The data display will show “80” with the second digit of 0 flashing.

f. Press the [š] key seven times. The data display will show “87” with the second digit of 7

flashing.

g. Press the [ENT] key. The message display will show “ PASS .”

(2) Setting Password 070

a. Press the [>] key once. The data display will show “000” with the second digit of 0 flashing.

b. Press the [š] key seven times. The data display will show “070” with the second digit of 7

flashing.

c. Press the [ENT] key. The message display will show “HIF.”

d. Press the [YES] key.

(3) Checking the date and time

a. Press the [ENT] key. The data display will show the date in day-month-year order.

b. Press the [ENT] key. The data display will show the time in hour-minute-second order.

c. Press the [ENT] key. The message display will show “mA.”

This is the inspection mode. When the message display shows “mA”, check the current outputs.
Each time the [ENT] key is pressed, the value on the data display increases in steps of 4.0 like

“4.0,” “8.0” and “12.0.” Check the current outputs. Corresponding to the data display, the

current output must be within the range shown in Table 1.
When the data display shows “12.0,” pressing the [ENT] key causes the message display to

show “RIPPLE” and the data display to remain unchanged. Press the [ENT] key again. The

message display will show “mA” and the data display “16.0.”

Table 1

Data Display

Current Output (mA)

4.0 4

±0.02

12.0 12

±0.02

20.0 20

±0.02

3.4 Temperature

Indication

Check

Following Section 3.3, press the [ENT] key until the message display shows “PT1000.” In this

state, change the resistance value of the decade resistance box as shown in Table 2 and check

the data display. The corresponding value on the data display must be within the range shown

in Table 2.