Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 20

Advertising
background image

IM 12B07D02-01E

Specification 2-7

2-4. Control Drawing of PH202S mA HART® Specification (IECEx).

PH202S (pH/O R P-transmitter)

Safe area

U o = 31.5 V o lt D C
Io = 100 mA
Po = 1.2 W att

EX ia or ib C ertified R epeater
Pow er Supply
(HAR T C ompatible)

O utput

Supply

SE N SO R (S)
term inals 11-17

+

_

G

SEN SO R (S)
term inals 11-17

Zo ne 0 o r 1

Hazardous area

Safe area

+

_

G

Functional
earth

Functional

earth

Intrinsically safe design
IEC -Ex standard EX ia IIC : T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

C ertificate nr. IEC Ex KEM 06.0052X

PH202S

+

_

Load

R esistance

EX ia or ib
C ertified safety barrier or pow er
w ith R int= 300

:

Io = 100 mA

U o = 31.5 V o lt D C

Zo ne 0 o r 1

Hazardous area

Functional
earth

+

_

24 volts D C N ominal

Supply V oltage.

Intrinsically safe design
IEC Ex standard EX ia IIC : T4 for ambient temp. < 55°C

T6 for ambient temp.< 40°C

C ertificate nr. IEC Ex KEM 06.0052X

(HAR T compatible)

(pH/O R P-transmitter)

Sensor(s) are of a passive type to be regarded as ‘simple apparatus’.

Electrical data of the PH202S.

- Supply and output circuit (terminals + and -):

Maximum input voltage U

i

= 31.5 V. Maximum input current I

i

= 100 mA.

Maximum input power P

i

= 1.2 W.

Effective internal capacitance

C

i

= 22 nF.

Effective internal inductance

L

i

= 35 PH.

- Sensor input circuit (terminals 11 through 17):

Maximum output voltage U

o

= 14.4 V. Maximum output current I

o

= 32.3 mA.

Maximum allowed external capacitance

Co = 600 nF. (for PH202S-E,-C,-U),

Co = 3.5 PF (for PH202S-N).

Maximum allowed external inductance

Lo = 34 mH (for PH202S-E,-C,-U),

Lo = 76 mH (for PH202S-N).

Barriers and power supply specification must not exceed the maximum values as shown

in the diagram above. These safety descriptions cover most of the commonly used

industry standard barriers, isolators and power supplies.

The Hand Held Communicator must be of a IECEx certified intrinsically safe type in

case it is used on the intrinsically safe circuit in the hazardous area or of a IECEx

certified nonincendive type in case it is used in the nonincendive circuit in the

hazardous area.