Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 22

Advertising
background image

IM 12B07D02-01E

Specification 2-9

2-6. Control Drawing of PH202S mA HART® Specification (FM Intrinsically safe design)

Sensor(s)
terminals 11-17
Max. cablelength: 60 mtr.
Cable dia. : 3…12 mm.

Sensor(s)
terminals 11-17
Max. cablelength: 60 mtr.
Cable dia.: 3…12 mm.

Ùnclassified Location

FM Approved

Power Supply

(HART compatible)

Output

Supply

+

_

G

Classified Location

Unclassified Location

+

_

G

Functional

earth

Functional
earth

Intrinsically safe design
FM Class I, Div.1, Group ABCD,

T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

PH202S transmitter

+

-

Load

Resistance

Classified Location

Functional
earth

+

_

24 volts DC Nominal

Supply Voltage.

FM Approved safety barrier or
power supply
with Rint = 300

:

(HART compatible)

For electrical data:
see text below.

For electrical data:
see text below.

Intrinsically safe design
FM Class I, Div.1, Group ABCD,

T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

PH202S transmitter

Figure 1

Figure 2

Electrical data of the PH202S.

-Supply circuit (terminals + and -):

Maximum input voltage Vmax = 31.5 V.

Maximum input current Imax = 100 mA.

Maximum input power Pmax = 1.2 W.

Effective internal capacitance Ci = 22 nF.

Effective internal inductance Li = 35 PH.

- Sensor input circuit (terminals 11 through 17):

Maximum output voltage Vt = 14.4 V. Maximum output current It = 32.3 mA.

Maximum allowed external capacitance Ca = 600 nF.

Maximum allowed external inductance La = 34 mH

If Hand Held Terminal (HHT) is not connected to the power supply lines of the PH202S

(see figure 1):

Any FM Approved barrier or power supply may be used that meets the following requirements.

Voc or Vt d 31.5 V ; Isc or It d 100 mA; Ca t 22nF + Ccable ; La t 35PH + Lcable

If HHT is connected to the power supply lines of the PH202S (see figure 2):

The Hand Held Terminal must be FM Approved. Refer to the manufacturers control drawing of

the HHT and the barrier/power supply to determine the cable parameters.

(Voc or Vt ) + VHHT d 31.5 V; (Isc or It ) + IHHT d 100 mA;

Ca t 22nF + Ccable+ CHHT ; La t 35PH + Lcable+ LHHT

When installing this equipment, follow the manufacturer’s installation drawing.

Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of Intrinsically Safe

Systems for Hazardous (Classified) Locations” and the National Electrical Code (ANSI/NFPA 70).

Control equipment connected to the barrier/power supply must not use or generate more than

250 Vrms or Vdc.

Resistance between Intrinsically Safe Ground and earth ground must be less than 1.0 Ohm.

In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.

WARNING

- Substitution of components may impair Intrinsic Safety

- To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing

or read, understand and adhere to the manufacturer’s’live maintenance procedures.

Application Doc. No.: IKE024-A10 P.4-1 to P.4-2