2. operating specifications, Caution – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 15

Advertising
background image

IM 12B07D02-01E

2-2 Specification

2-2. Operating specifications

A. Performance : pH

- Linearity :

±0.01 pH

- Repeatability:

±0.01 pH

- Accuracy :

±0.01 pH

Performance :

ORP

- Linearity :

±1 mV

- Repeatability :

±1 mV

- Accuracy :

±1 mV

Performance :

Temperature with Pt1000 Ω,

3kΩ Balco, 5k1Ω, 350Ω, 6k8Ω, PTC10kΩ &

8k55Ω

- Linearity :

±0.3 ˚C

- Repeatability:

±0.1 ˚C

- Accuracy :

±0.3 ˚C

Performance :

Temperature with Pt100 Ω

- Linearity :

±0.4 ˚C

- Repeatability:

±0.1 ˚C

- Accuracy :

±0.4 ˚C

Note on performance specifications:

The following tolerance is added to above

performance.

mA output tolerance : ± 0.02 mA of

"4 - 20 mA"

B. Ambient operating temperature

:

-10 to + 55 ˚C (10 to 131 ˚F)

C. Storage temperature

: -30 to +70 ˚C (-20 to 160 ˚F)

D. Humidity

: 10 to 90% RH (Non-condensing)

F. Housing:

Case: Cast aluminum case with chemically

resistant coating

Cover: Polycarbonate window.

Case color: Off-white (Equivalent to Munsell

2.5Y8.4/1.2)

Cover color: Deep sea Moss green

(Equivalent to Munsell 0.6GY3.1/2.0)

Grounding

terminal

Power/Output

cable gland

Sensor

cable gland

Cable and terminals :

The PH202 is equipped with terminals

suitable for the connection of finished

cables in the size: 0.13 to 2.5 mm (26 to

14 AWG)

Cable entry: 2 cable glands 1/2NPT. The cable

glands will form a tight seal on cables with

an outside diameter in the range of 6 to 12

mm (0.24 to 0.47 inches).

Construction : Weather resistant to IP65 and

NEMA 4X standards

Mounting : Pipe, wall or panel mounting, using

optional

hardware

Weight : Approx. 1.6 kg

G. Shipping details

: Package size w x h x d

290 x 300 x 290 mm.

11.5 x 11.8 x 11.5 in.

Packed weight approx. 2.5 kg (5lb)

H. Data protection

: EEPROM for configuration and logbook, and

lithium cell for clock.

I. Watchdog timer

: Checks microprocessor

J. Automatic safeguard

: Return to measuring mode when no keystroke

is made for 10 min.

K. Operation protection

: 3-digit programmable password.

L. Sensor impedance checking

: Independent impedance check on measuring

and reference sensor elements, with

temperature compensation. Display of sensor

impedance on message line of display. FAIL

flag in event of “out of limits” impedance, and

the possibility of 21 mA or (3.6 mA or 3.9 mA)

error signal.

M. Signal processing (pH/ORP)

: The PH202 can measure pH or ORP. Using

the FU20 allows simultaneous measurement

and display of pH and ORP. It also allows

display and out put of pH.

N. EMC Conformity standards

,

EN 61326-1 Class A, Table 2

(For use in industrial locations)

EN

61326-2-3

EN 61326-2-5 (pending)

CAUTION

This instrument is a Class A product, and it is

designed for use in the industrial environment.

Please use this instrument in the industrial

environment only.

Advertising