2. configuration checklist for ph202g, 2. configuration checklist for ph202g -3, Ma ma ma ma – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 89

Advertising
background image

IM 12B07D02-01E

Appendix 10-3

10-2. Configuration checklist for PH202G

Standard Configuration

Options

Reference for
change

Measured Variable(s)

primary inputs

pH, ORP and Temp

pH range

0-14 pH

any span within -2-16 pH

“output”

pH range linearized

disabled

21 point table

codes 31& 35

ORP range

-500 to 500 mV

spans up to 3000 mV between

“output”

-1500 to 1500mV

Temperature range

-30-140°C

Temperature unit

Celsius

Fahrenheit

code 11

mA Outputs

analog output

4- 20 mA for pH

pH/ORP/(parameter 2)

code 01, 02, 31

output linearization

disabled

pH/ORP

codes 35

Communication

digital interface

disabled

HART

60

communication software

external

HHC or PC202

contact factory

variables on display

pH/ORP and temp

pH/ORP, parameter 2, mA output

“display”

SL, AP, Z1, Z2 etc.

burn out

disabled

burn low (3.6 or 3.9)/ high (21) on mA output code 32

password protection

disabled

for maint/ comm./ serv level

code 52

autoreturn

return to measure in 10 min. enable or disable

code 50

add. functions in MAINT

disabled

Impedance check start

code 51

Diagnostics

impedance checking

active

enable or disable

code 03 & 04

check on calibration data

active

enable or disable

code 05

check on stability

0.02 pH per 5 s

choose stability level

code 20

display calibration log.

enabled with logbook

diagnostics setup

codes 03, 04 & 05

Compatibility

pH or ORP

glass sensor/metal electrode pH or ORP

code 01

temperature sensor

Pt 1000Ω

Pt1000; Pt100, etc

code 10

other sensors

enamel sensors (Pfaudler)

ITP & impedance check setup

codes 23, 03 & 04

2nd parameter

disabled

pH & ORP/ pH &rH

code 02

manual temp. comp.

disabled

disable or enable

“temp”

Special Features

buffer table configuration

NIST standard

fully configurable

codes 24, 25 & 26

temperature calibration

none

adjustment +/- 20 °C

code 12

zero point calibration

disabled

disable or enable

code 27

call for maintenance

set time interval 1 - 250 days

code 55

HOLD during maintenance

hold last or hold fix

“hold”

process temp. compensation disabled

set temperature coefficient

code 13

logbook

disabled

2 x 50 events

code 61, 62

mA
mA
mA
mA