Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 115

Advertising
background image

IM 12B07D02-01E

Appendix 12-3

3/3

QIS 12B07D02-01E

Table 3

Check Point (pH)

Simulation Input (mV)

Data Display (pH)

0 +414.1

0.00

±0.01

7 0.0

7.00

±0.01

14 -414.1

14.00

±0.01

3.5 ORP

Indication

Check

Following Section 3.5, press the [ENT] key. The message display shows “ORP.INP.” In this state,

change the simulation input as shown in Table 4 by means of the standard voltage/current source

and check the data display. The corresponding value on the data display must be within the

range shown in Table 4.

Table 4

Simulation Input (mV)

Data Display (mV)

-1500

-1500

±1

0 0

±1

+1500

+1500

±1

Press the [ENT] key until the message display shows “READY.”

Press the [ENT] key again to end the tests.

Figure 1 Testing Circuit and Test Equipment

DC source

24 V DC

Standard

Voltage Source

+

300 :

+

-

-

DC

Milli-

ammeter

Ground

-

+

Shorting bar

Decade

Resistance

Box

PH202G, PH202SJ

SUPPLY

TEMP

INPUT 2

INPUT 1

LE

LOW IMP HIGH IMP

+

-

G 11 12

14

17

13

15

16

Advertising