Appendix, 1. user setting table, Appendix -1 – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 87: 1. user setting table -1

Advertising
background image

IM 12B07D02-01E

Appendix 10-1

10. APPENDIX

10-1. User setting table

FUNCTION

SETTING DEFAULTS

USER SETTINGS

Parameter specific functions

01 *PH.ORP 0

pH

02 *PRM2 0

Off

03 *Z1.CHK

1.1.1

High range, TC on

check

on,

04 *Z2.CHK

0.0.1

Low range, TC off

check off no TC

05 *CAL.CK

1.1

AP on, Slope on

Temperature functions

10 *T.SENS 0

Pt1000

11 *T.UNIT 0

°C

12 *T.ADJ

None

13 *T.COMP 0

Off

*T.COEF

-0.00 pH/10°C

Calibration functions

20 *∆t.SEC 5

Sec

*∆PH 0.02 pH

21 *AS.LOW -120

mV

*AS.HI 120

mV

22 *SL.LOW 70

%

*SL.HI 110

%

23 *ITP

7.00

pH

*SLOPE 100.0 %

*ASP.1D

0.0

mV

*ASP.mV

mV

24 *BUF.ID 4

NIST

4

25 *BUF.ID 7

NIST

7

26 *BUF.ID 9

NIST

9

27 *ZERO.P 0

disabled

mA outputs

31 *OUTP.F 0

pH

(ORP)

32 *BURN 0

off

35 *TABLE

21 pt table see code 31

mA

mA
mA

mA
mA

mA