Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6020-14

1316A-1 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 59 of 98

Figure 4.3. The Status Window displaying Zero calibration in

Progress.

The Zero/O

2

Span is started controlled by the BZ6012 going through

the following steps:

i) The pump is turned OFF and all solenoid valves are deactivated.
ii)

After a settling delay, the ambient pressure (Atm Press) is

measured and stored.

iii) The pump is turned ON.
iv) After a settling delay, the sample cell pressure (Abs Press) is

measured and stored.

v)

The absolute pressure limit is checked (Abs Press – Atm Press)

Pressure Limit.

If this test fails, the Pump Out Flow Check is shown in the Status

window (see

figure 4.4

) and the zero calibration procedure is aborted: