Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6020-14

1316A-1 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 60 of 98

Figure 4.4. The Status Window displaying the Pump Out Flow Check.

The Multi Gas Monitor – INNOVA 1316A-1 is purged with sample gas

by pumping room air from the zero inlet on the rear panel through the

entire flow system.

During the zero calibration purge period, the O

2

sensor’s electrical

zero offset is calculated and stored. The optimum gain is calculated

for each IR channel.
The new calibration constants:
CO

2

, CO, HC, and NO zero calibration constants.

O

2

span calibration constant.

Are calculated and stored in flash memory:
When the zero calibration process is successful, the following actions

are performed:
The red Zero Required in the Status screen, see

figure 4.3

, is

changed to show a green Not Required, see

figure 4.5

.