Figure 4.14 – LumaSense Technologies INNOVA 1316A-1 User Manual

Page 70

Advertising
background image

Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6020-14

1316A-1 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 70 of 98

Figure 4.14 The Status window displaying that a new O

2

sensor is required.

Prior to any span calibration perform a Zero/O

2

span calibration after

completion of the warm up time.

Initially set Zero Calibration to "Manual" in “Zero Group”, see figure 4.15.

This will avoid the Zero calibration to interfere with the Span calibration.

Figure 4.15 Set Zero calibration to Manual

Start the measurement by pressing the “Start”-button in the Main window, see

figure 4.16