1316A-1 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 80 of 98

1. Use to tick which of the CO, CO

2

, CHx gasses that you want

to be calibrated.

2. Type in the correct concentrations for the ticked CO, CO

2

, CHx

gas.

The gas concentration must be the same as that stated on the

calibration gas bottle.

3. Connect the calibration-gas bottle as described in

section

4.3.4

.

4. Open the gas valve and adjust the flow so that a small surplus

flow exists in the side-stream tube.

5. Press the Send button to start the gain calibration.

After making the Span Calibration please make sure that the Mode is

set back into 5 and that the Zero calibration is set to “Auto”, as

described in

section 4.3.5

.

The monitor is now ready for making measurements.

Press Send.

The Process State In Progress is displayed in the Span Calibration

window (see

figure 4.30

) and the Data Status in the Status window is

changed to show HC Span, see

figure 4.31

.

The 1316A-1 is calibrated with the selected gas at the concentration

keyed in.