Tick the gas(ses) to be Span calibrated, and key in concentration value(s), see

figure 4.23. Press the “Send” button to perform the Span calibration. Observe

the “Process State” change to “In Progress” during the Span calibration, see

figure 4.24

.

Figure 4.23 The Span Calibration Window to select the gasses and the

concentrations for Span calibration

The text box in the Span Calibration window now displays Span Calibration ACK.

The Span Calibration is in progress.