5 Press the Send button to start the gain calibration.

After completion of the span calibration routine:

The validity of each new span calibration constant (CO

2

, CO, HC)

is checked.
New span calibration constants for the HC, CO and CO

2

channels

should not significantly differ from the factory span calibration

constants.
If a newly calculated span calibration constant does differ by

greater than ±30% from the original factory value, the

appropriate span fail status (CO

2

Span Fail, CO Span Fail, HC

Span Fail), is set true see figure 4.34.:

Figure 4.34. Status window showing a CO2 span failed.

If the NO channel is being span calibrated, the current full-scale NO

output is calculated and saved. If the current full-scale NO output is less

than or equal to a preset value (The Low NO Sensor Limit) the New NO

Sensor Required is shown in the Status window, see

figure 4.13

.