Rt‐15 flow monitor – AW Gear Meters RT-15 User Manual
Page 16
RT‐15 Flow Monitor
Operation and Programming Manual
15
Startup Message
The RT‐15 displays an initial window at power up. After one second, the following message displays in
the window for approximately two seconds. After this message, the display reverts to the Display Mode
you last selected.
LOAD VARS
– Indicates that the variables are being recalled from memory.
TURN ON
– Increments on each power‐on cycle.
MEM LEFT
– Memory life indicator: displays percent of memory life remaining in .1% increments.
The RT‐15 stores programmed variables and retained totalizer values in EEPROM (electrically erasable
programmable read‐only memory). The limit on the number of times memory can be re‐written results in
a minimum EEPROM life of five years with continuous use. You can replace this memory IC if necessary
(contact AW Gear Meters). You must re‐program variables after a memory IC (integrated circuit) change.
VERSION
– Indicates the software version installed in the RT‐15.
Display Modes
The RT‐15 has ten display/program modes:
LOGO
RATE A
GT
RATE A
PL
TOTAL A
GR TOTAL A*
STATUS ONE*
ANALOG OUT
LIMITS
RAT A TOT A*
BATCH TOTAL
*Display only; no programming function.
LOAD VARS
RT‐15
TURN ON
0026.
MEM LEFT 099.9
VERSION 222 2006
Notice