RT-30EX Flow Transmitter

Operation and Programming Manual

29

RT-30EX User Manual, Rev. 1.7 02/07/2012

cleared. If this is correct, press the

YES

button. This will cause all table values to be set to zero and the

Linearizer will be turned OFF. The user has the ability to press

NO

button to cancel this option and

return to Point 1 programming screen.

Changing the linearizer table

To change a table Point value, enter the programming screen and press the

ENT

button to get to the

table value to change. After changing the relevant values, continue pressing the

ENT

button through the

remaining table values until reaching the end of the programmed table (first table point whose Freq. is
zero) to exit the table.

Adding linearizer table point

To add more table points, it must be noted that data can only be added to the end of a table. Therefore,
if the additional point(s) to add do not have a frequency greater than the last entered point, it will
require the user to manually “shift” the table points by entering the new points after the next smallest
existing value and then re-entering the remaining points.

If an existing table is changed and has more points than required, once the required points have been
entered and the next table value has the Freq. changed to zero, all remaining points will automatically
be reset to zero when exiting the table.