Fine limit output adjustment, Coarse limit output adjustment, On/off limit output adjustment – AW Gear Meters RT-30EX User Manual

Page 33: Rt-30ex flow transmitter

Advertising
background image

RT-30EX Flow Transmitter

Operation and Programming Manual

33

RT-30EX User Manual, Rev. 1.7 02/07/2012

> Fine adjustment
Coarse adjustment
ON/OFF adjustment

Enter frequency value
to output

Output = 0500.00 Hz

SEL

ENT

Fine limit output adjustment

When choosing the Fine adjustment option, the screen will appear which shows the active frequency
output value on the Limit 1, Limit 2 and Frequency outputs. Change this number to any desired value
between 2Hz and 4,500Hz and when

ENT

button is pressed, the RT-30EX will output this value on the 3

outputs at the same time. This also returns the screen to the adjustment type screen. Press

SEL

to see

output value again and make new change.

Coarse limit output adjustment

When choosing the Coarse adjustment option, the screen will appear where the frequency output value
can be changed. When pressing the

UP

button, the screen value (and Hz output value) increments by

25Hz to a max of 4,500Hz. When pressing the

DN

button, the screen value (and Hz output value)

decrements by 25Hz to a min of 2Hz. Press the

EXIT

button to leave this mode.

ON/OFF Limit output adjustment

When choosing the ON/OFF adjustment option, the screen will appear showing the current state of each
output pin. By pressing the

UP

,

DN

and

SEL

buttons, each associated output as shown in Figure 45

above will change the state.

Use UP/DN keys to
change output in
25 Hz increments

Output = 0500.00 Hz

Fine adjustment
> Coarse adjustment
ON/OFF adjustment

SEL

ENT

Fine adjustment
Coarse adjustment
> ON/OFF adjustment

UP > LIMIT 1 = OFF
DN > LIMIT 2 = OFF
SEL > FREQUENCY = OFF

SEL

EXIT

Figure 43: Fine Limit adjustment screens

Figure 44: Coarse Limit adjustment screens

Figure 45: ON/OFF Limit adjustment screens