For example:

A flowmeter has a K-factor 2053.7 imp/Gal and the display should read in Gal/Min.

KFR = 6000 / 2053.57 = 2.921741 for GPM

292174 is > 65535! 29217 > 4 digits! 2922 is best! Enter 2922 for KFR. For this

example, KFR=2.922, the decimal point is set to three decimal places in the next procedure.


Press and hold the (UP) or (DN)-key to increment or decrement the variable value. To increase
the increment or decrement rate, while holding the (UP) or (DN)-key, press and hold the (SEL)-
key for medium rate or the (ENT)-key for maximum rate. When the desired value is displayed,
press the (ENT)-key to store the programmed value and advance to the next variable.

NOTE: Always release (SEL) or (ENT)-key before releasing the (UP) or (DN)-key or the LLC
will proceed to next variable leaving the last variable with an unpredictable value.


DP for KFR
- This variable selects the location of the decimal point in the KFR factor and the
display. This display will always be zeros with a decimal point.

Factory default is

00000.00


Use the (UP) or (DN)-key to move the decimal point left or right. Press the (ENT)-key to store the
location and advance to the next variable.


Eng

=g UNITS KFR - The desired engineering units for display of rate are entered using up to three

programmable characters or letters. Example: gpm

Factory default is

hz_


The first (left-most) character is programmed/edited first. To choose a letter press the (UP) or (DN)-
key until the desired character or letter appears. Press the (SEL)-key to enter the 1st character. Repeat
the procedure for the 2nd character and enter with the (SEL)-key. Repeat for the 3rd character and
store the last entry with the (ENT)-key. The next variable will display.