7.4 GROUND RESISTIVITY MEASUREMENT
For the ground resistivity measurement the test current is injected through two current
probes (C1/H and C2/E). The voltage probes S and ES must be placed between the
current probes (equidistance ‘a’ between probes must be considered). Using different
distances between the test probes means that the material is measured at different
depths. By increasing the distance ‘a’ a deeper layer of ground material is measured (see
Fig. 27)

Fig. 27: Connection of the unit for ground resistance measurement

1 Use

or

keys and select the EARTH RE measurement on the station unit

2 Use the  or  keys to select the “EARTH

” measurement. The below screen is

displayed:

Fig. 28: Main screen of ground resistance measurement


3 Use

the

 or  keys to select the “a” parameter. Use the  or  keys to set the value

expressed in meters into the range 0.1 ÷ 30.0 by steps of 0.1m

4 Fit the auxiliary test probes at the same distance “a” set on the station unit and connect

the input terminals C1/H, C2/E (current probes), ES and S (voltage probes) to the earth
system under test by using the cables supplied as accessories (see Fig. 27)

5 Press

TEST

key to perform the measurement. The result is displayed

Fig. 29: Result screen of ground resistance measurement

Where:

  Measured value of ground resistivity
Rp  Voltage probe resistance (sum of S+ES)
Rc  Current probe resistance (sum of C1/H +C2/E)