HT4020 - HT4022

EN - 15

4.3.8. Power/Energy

measurement on single phase system

CAUTION

The maximum Voltage input is 600V. Do not attempt to take any voltage
measurements exceeding the limits indicated in this manual. Exceeding the
limits could cause electrical shock or damage to the instrument.

Fig. 11: Use of instrument for power/energy measurement on single phase system

1. Select the “W” position
2. Open the jaws and clamp only one cable. Pay attention to the alignment marks (see §

4.1.2. and Fig. 11)

3. Insert the red plug into V/

 jack and the black plug into COM one

4. Connect the test leads to the circuit under test (see Fig. 11). The measured active

power value is displayed

5. If the symbol "

" is displayed the input voltage and/or current value is higher than the

instrument’s full scale. Therefore the power and power factor values could be incorrect

6. By pressing the

FUNC key the parameters active power (kW), reactive power (kVA

R

,

capacitive C, inductive I), apparent power (kVA), power factor (Pfi or Pfc for inductive
and capacitive respectively) are displayed

7. Press and hold the ENERGY key for at least 1 second to active the energy

measurement mode. By pressing the

FUNC key the parameters active energy

(kWh), reactive energy (kVA

R

h, capacitive C, inductive I), apparent energy (kVAh),

TIME with indication of energy measurement duration are displayed

8. Press the ENERGY key to activate the energy measurement. The message

MEASURING” appears at the bottom of the display. Press the ENERGY key again to
stop the energy measurement, the message “MEASURING“ disappears from the
display

9. For HOLD and MAX/MIN/AVG features refer to § 4.2.1 and § 4.2.3
10. Press ENERGY key for at least 1 second to escape this mode going back to power

measurement function