Technical specifications, En - 23 – HT instruments HT4022 User Manual

Page 56

Advertising
background image

HT4020 - HT4022

EN - 23

6. TECHNICAL SPECIFICATIONS

6.1. TECHNICAL

CHARACTERISTICS

Accuracy indicated as [%rdg + (num. digit * resolution)] referred to: 23°C±5°C, <75%RH

DC Voltage

Range Resolution

Accuracy

Input

impedance

0

 599.9V

0.1V

(1.0%rdg + 3dgt)

1M

AC TRMS Voltage

Range Resolution

Accuracy

Input impedance

40

200Hz

200

400Hz

1.6

 599.9V

0.1V

(1.0% rdg+3dgt)

(5.0%rdg + 3dgt)

1M

Max. Crest factor = 1.41

MAX / MIN / AVG / PEAK AC/DC Voltage

Function Range Resolution

Accuracy

Response

time

MAX,MIN,AVG

10.0

 599.9V

0.1V

(5.0%rdg + 10dgt)

500ms

PEAK

10

 850V

1V

1ms

AC TRMS Current

Range Resolution

Accuracy

Overload protection

40

200Hz

200

400Hz

0.0

 399.9A

0.1A

(1.0%rdg+3dgt)

(5.0%rdg + 5dgt)

600A RMS

Max. Crest factor = 2

MAX / MIN / AVG / PEAK AC Current

Function Range Resolution

Accuracy

Response

time

MAX,MIN,AVG

10.0

 399.9A

0.1A

(5.0%rdg + 10dgt)

500ms

PEAK

10

 800A

1A

15ms

Resistance and Continuity test

Range Resolution

Accuracy

Overload

protection

0.0

 499.9 0.1

(1.0%rdg + 5dgt)

600V AC/DC RMS

500

 999

1

1000

 1999

3

Instrument emits a buzzer for R<40

Frequency (with test leads/ with jaws)

Range Resolution

Accuracy

Overload

protection

40.0

 399.9Hz

0.1Hz

(0.5%rdg + 1dgt)

600V RMS / 600A RMS

Voltage range for frequency measure, with test leads: 0.5

 600V, with jaws: 0.5  400V

Voltage and current harmonics (HT4022 only)

Harmonic order

Resolution [V], [A]

Accuracy

1

 15

0.1

(10.0%rdg + 5dgt)

16

 25

0.1

(15.0%rdg + 5dgt)

Accuracy defined for: voltage

1.6V, current 2A

Power factor

Range Resolution

Accuracy

0.20

 1.00

0.01

 3°

Accuracy defined for: sine wave, voltage 230

 400V, current 2A, freq. 50-60Hz

Active power/energy, Reactive power/energy, Apparent power/energy

Range

[kW,kWh], [kVAR,kVARh], [kVA, kVAh]

Resolution

[kW,kWh], [kVAR,kVARh], [kVA, kVAh]

Accuracy

0.00

 99.99

0.01

(3.5%rdg + 3dgt)

100.0

 999.9

0.1

Accuracy defined for: sine wave, voltage 230

 400V, current 1A, freq. 50-60Hz, Pf: 0.8i  0.8c

Phase sequence and phase coincidence

Range

Frequency range

Input impedance

Overload protection

80

 600V

40

 69Hz

1M

600V RMS

Advertising
This manual is related to the following products: