5.5.1.2 Entering Known Calibration Curve Constants

The Use Standard Samples checkbox should be

clear and the Curve Fit option should be set to

the appropriate option. The known calibration

curve constants should then be entered into the

appropriate K0 – K3 fields where:-

Concentration = (K3*X

3

)+ (K2*X

2

)+ (K1*X) + K0

X = Photometric value of the sample

5.5.2

Constructing a New Calibration Curve

To construct a new calibration curve the user

must ensure that they check the tick boxes beside

the Calculate Concentration and Use Standard

Sample options.

5.5.2.1 Measuring Standard Samples

Select the Standard tab.

Insert the first standard into the sample position

in the sample chamber and the reference into the

reference position in the sample chamber.

Select the Start button to initiate the

measurement

A new window will open allowing the Standard

name, final photometric reading and

concentration to be edited. The actual

photometric reading of the standard will be

displayed in the Readout window and this value

will default to the deltaAbs field.