SECTION 9 – APPENDIX

The multicell accessory measurement mode allows the user to measure a sample’s photometric

absorbance or %transmittance at up to 10 wavelengths. Following sample measurement, the

photometric readings are displayed on the multi-wavelength mode screen.

9.1 CALCULATIONS IN QUANTITATION MODE

Single Wavelength Method

:Abs.=A

1

Double Wavelengths Method

:Abs.=m*A

1

-n*A

2

Three Wavelengths Method

:Abs.=A

1

-(WL1-WL2)*(A

2

-A

3

)/(WL2-WL3)-A

3

9.2 CALCULATIONS IN DNA/PROTEIN MODE

260/280:

CDNA

=(A1-Aref)*f1-(A2-Aref)*f2

CProtein =(A2-Aref)*f3-(A1-Aref)*f4
Ratio

=(A1-Aref)/(A2-Aref)

A1=A260nm, A2=A280nm, Aref=A320nm (Optional)

f1=62.9,

f2=36.0,

f3=1550,

f4=760.0


260/230:

CDNA

=(A1-Aref)*f1-(A2-Aref)*f2

CProtein =(A2-Aref)*f3-(A1-Aref)*f4
Ratio

=(A1-Aref)/(A2-Aref)

A1=A260nm, A2=A230nm, Aref=A320nm (Optional)

f1=49.1,

f2=3.48,

f3=183,

f4=75.8