8.4.2

Constructing a New Calibration Curve

8.4.2.1 Measuring Standard Samples

Select the Standard tab.

Insert the first standard into the sample position in

the sample chamber and the reference into the

reference position in the sample chamber.

Select the Start button to initiate the measurement

A new window will open allowing the Standard

name and concentration to be edited. The actual

photometric reading of the standard will be

displayed in the Readout window. Select OK to

transfer the information to the main standard tab.

Repeat this procedure for each standard sample.

8.4.3 Sample

Measurement

To quantify an unknown sample select the

Sample tab and select the Start button.