Getting started – KYORITSU 4202 User Manual

Page 11

Advertising
background image

̶ 9 ̶

 6. Getting started

(1) Check the battery voltage

  When nothing is displayed on the LCD, press the power button

and turn

on the instrument. Battery voltage is enough when indication is clear and the

"

" symbol is not displayed on the LCD after turning on the instrument.

  Follow the procedure described in “9. Battery replacement” and replace the

batteries with new ones when any of following symptoms is noted. Otherwise,

accurate measurement and proper saving cannot be ensured.

  * "

" symbol is being displayed.

  * indications are faint and difficult to read.

  * nothing is displayed on the LCD.

(2) Verify the correct measurement of earth resistance

  Clamp-on the supplied resistor for operation check as shown below, and verify

that the Transformer jaw and the circuit works correctly. When the indicated

value is within the range described below, they are operating correctly. If the

indicated value is greatly exceeding the accuracy, send the instrument to your

local KYORITSU distributor for repair or recalibration.

  For repair, the resistor for operation check shall be attached and returned

together with the instrument.

#

CAUTION

This instrument performs self-calibration for about 3 seconds when it is turned

on. ("

" is displayed on the LCD.) Do not clamp onto any conductor or

open the jaws in this period. Otherwise, inaccurate measurement may occur.

Resistor for operation check

1 Ω

10 Ω

Allowable range

0.93 ∼ 1.07

9.75 ∼ 10.25

Advertising