KYORITSU 4202 User Manual

Page 7

Advertising
background image

̶ 5 ̶

Auto power-off

Turns power off about 10 minutes after the last

button operation.

Applicable standards

IEC61010-1 (CAT. IV 300V Pollution degree2)

IEC61010-2-032

IEC61326-2-2 (EMC standard)

External communication Bluetooth Ver2.1+EDR Class2

method

Withstand voltage

AC5160Vrms/ 5 seconds

Between the Transformer jaws fitted parts and Case

enclosure (except for jaws)

Insulation resistance

50MΩ or more at 1000V

Between the Transformer jaws fitted parts and Case

enclosure (except for jaws)

Conductor size

Approx. 32mm in diameter max.

Dimension

246(L) x 120(W) x 54(D)mm

Weight

Approx. 780g (including batteries)

Accessories

Battery R6P: 4pcs (MODEL4200)

LR6: 4pcs (KEW4202)

Instruction manual

: 1pce

Resistor for operation check : 1pce

(MODEL8304)

Hard case MODEL9166

: 1pce (MODEL4200)

MODEL9167

: 1pce (KEW4202)

<Supplemental remarks>

○ Effective value (RMS)

Most alternating currents and voltages are expressed in effective values, which are also

referred to as RMS (Root-Mean-Square) values. The effective value is the square root of

the average of square of alternating current or voltage values. Many clamp meters using

a conventional rectifying circuit have "RMS" scales for AC measurement. The scales are,

however, actually calibrated in terms of the effective value of a sine wave though the clamp

meter is responding to the average value. The calibration is done with a conversion factor of

1.111 for sine wave, which is found by dividing the effective value by the average value.

These instruments are therefore in error if the input voltage or current has some other shape

than sine wave.

○ CF (Crest Factor) is found by dividing the peak value by the effective value.

  Examples:

  Sine wave: CF=1.414  Square wave with a 1: 9 duty ratio: CF=3

Advertising