Specification – KYORITSU 4202 User Manual

Page 6

Advertising
background image

̶ 4 ̶

 3. Specification

Operating system

Earth resistance function: Constant voltage injection,

Current detection,

(Frequency: Approx.2400Hz)

Dual Integration

AC current function:Successive Approximation(True-RMS)

Display

Liquid crystal display with a maximum count of 2099

Over-range indication

"OL" is displayed when input exceeds the upper limit

of the measuring range

Response time

Earth resistance function :Approx. 7 seconds

AC current function :Approx. 2 seconds

Sample rate

Approx. once per second

Location for use

Altitude 2000m or less, Indoor/ outdoor use

IP protection degree

IP40

Temperature & humidity range 23℃ ±5℃ /Relative humidity 85% or less

(guaranteed accuracy) (no condensation)

Operating temperature & -10℃ to 40℃ /Relative humidity 85% or less

humidity range

(no condensation)

Storage temperature & -20℃ to 60℃ /Relative humidity 85% or less

humidity range

(without batteries, no condensation)

Power source

DC6V: R6P (size AA manganese battery) x 4pcs, or

LR6 (size AA alkaline battery) x 4pcs

Current consumption

MODEL4200 : Approx. 50mA (max. 100mA)

KEW4202 : Approx. 90mA (max. 140mA)

Measurement time

MODEL4200 : Approx. 12 hours (when using R6P), or

24 hours (when using LR6)

KEW4202 : Approx. 5 hours (when using R6P), or

21 hours (when using LR6)

Measuring range and accuracy

Function

Range

Resolution

Measuring range

Accuracy

Earth resistance

(Auto-ranging)

20Ω

0.01Ω

0.00 - 20.00Ω

±1%±5dgt

200Ω

0.1Ω

20.0 ‒ 200.0Ω

±3%±5dgt

600Ω

200 - 399Ω

±6%±5dgt

400 - 599Ω ±10%±10dgt

1200Ω

10Ω

600 - 1260Ω

-

AC current (ACA)

(50Hz/60Hz)

(Auto-ranging)

100mA

0.1mA

0.0 - 100.0mA

±2%±5dgt

(sine wave)

1000mA

1mA

100 - 1000mA

10A

0.01A

1.00 - 10.00A

30A

0.1A

10.0 - 30.0A

* Crest factor ≦2.5 Accuracy at sine wave +1% (50Hz/60Hz, peak value shall not exceed 60A)

* In the following cases, zero will be displayed on the LCD.

- At 20Ω range of Earth resistance function: 0.04Ω or less

- At 100mA range of ACA function: 0.4mA or less

* A range shifts to upper range when the input exceeds 105% of the selected range, and shifts

to the lower range when the input falls under 80% of the range.