3 measurement of earth resistance – KYORITSU 4202 User Manual

Page 14

Advertising
background image

̶ 12 ̶

#

CAUTION

Follow the procedure described in "7-1 Normal measurement of

current" and measure the current flowing on the earthed wire prior to the

measurement of earth resistance.

  In case that the "

" symbol is displayed at the upper right of the

LCD, it means that a great error would be included in the measured result.

To avoid such inaccurate measurement, reduce the current flowing on the

earthed wire by turning off the device from which current is applied to the

earthed line under test.

Measurement cannot be made for the earth without multi-earthing system or

when the earth resistance under test is smaller than the other earth resistances.

To avoid inaccurate reading may be taken, never make a measurement for

the same earth system with multiple units of this instruments.

The "

" symbol may be displayed during a measurement of earth

resistance. It indicates that the jaws of the instrument are not properly

closed. Measurement is being stopped while this symbol is displayed on the

LCD. Close the Transformer jaws properly to re-start the measurement.

Response time at Earth resistance function is about 7 seconds. It takes

awhile until the readings become stable.

7-3 Measurement of earth resistance

Measurement procedure

 * Press the Function button

and select the Earth resistance function.

 * Confirm the displayed unit is " Ω ", and " MEM " is not displayed at the upper

left on the LCD.

 * Press the trigger to open the Transformer jaws, and close them over the

earthed wire under test.

 * Measured resistance value is displayed on the LCD.

<Noise check function>

  At the Earth resistance function, the "

" symbol is displayed on the LCD in

the following cases which may effect on a measurement.

 * The current flowing on the earthed wire is exceeding the following value.

Range of Earth resistance function

Allowable current value

20Ω

2A or less

200Ω/600Ω/1200Ω

400mA or less

 * The current flowing on the earthed wire includes a harmonic wave which

effects on the measurement.

<Jaws check function>

  The "

" symbol is displayed when the Transformer jaws of the instrument are

not properly closed.

 Measurement will be stopped when this symbol appears on the LCD.