7-1 Normal measurement of current

  * Press the Function button

and select the ACA function.

  * Confirm the displayed unit is " mA ", and the " MEM " is not displayed at the

upper left on the LCD.

  * Press the trigger to open the Transformer jaws, and close them over one

conductor only.

  * Measured current value is displayed on the LCD.

  (Earth leakage current that flows through an earthed wire can be measured by

this method.)

7-2 Measurement of balance leakage current

  * Press the Function button

and select the ACA function.

  * Confirm the displayed unit is " mA ", and the " MEM " is not displayed at the

upper left on the LCD.

  * Clamp onto all conductors except for the earthed wire.

  * Measured current value is displayed on the LCD.

Load

Load

Single-phase 2-wire system

In 3-wire system with neutral,

clamp onto all 3 wires.

Three-phase 3-wire system

In 4-wire system with neutral,

clamp onto all 4 wires.