2 output – Pyromation Series 642 User Manual

Page 34

Advertising
background image

Technical data

Series 642

34

Pyromation, Inc.

9.0.2

Output

Output signal

Analog 4 to 20 mA, 20 to 4 mA

Signal on alarm

• Underranging:

Linear drop to 3.8 mA

• Overranging:

Linear rise to 20.5 mA

• Sensor break; sensor short-circuit (not for thermocouples TC):

≤ 3.6 mA or ≥ 21.0 mA (configurable 21.6 mA to 23 mA)

Load

Max. (V

power supply

- 11 V) / 0.022 A (current output)

Linearisation/transmission
behaviour

Temperature linear, resistance linear, voltage linear

Filter

1

st

order digital filter: 0 to 60 s

Galvanic isolation

U = 2 kV AC (input/output)

Input current required

≤ 3.5 mA

Current limit

≤ 23 mA

Switch-on delay

4 s (during switch-on operation I

a

= 4 mA)

Voltage transmitter (mV)

Millivolt transmitter (mV)

-20 to 100 mV

5 mV

1)

Increasing inaccuracy for temperatures < 300 °C (< 572 °F)

Input

Designation

Measuring range limits

Min. span