4 wiring, 1 quick wiring guide, Wiring – Pyromation Series 642 User Manual

Page 9: Quick wiring guide, 4wiring

Advertising
background image

Series 642

Wiring

Pyromation, Inc.

9

4

Wiring

NOTICE

Installation in hazardous area
► When installing Ex-approved devices in a hazardous area please take special note of the

instructions and connection schematics in the respective Ex documentation added to this
operating manual. The local representative is available for assistance if required.

For wiring the device proceed as follows:

1.

Remove the cover clamp (

→ å 3, item 1).

2.

Remove the device cover (

→ å 3, item 2).

3.

Remove the display from the electronics module (

→ å 3, item 3).

4.

Open the 2 screws of the electronics unit and remove the electronics unit (

→ å 3, item 4).

5.

Open the cable gland at the device (

→ å 3, item 5).

6.

Feed the cable through the opening in the cable gland.

7.

Connect the wires (

→ å 5).

8.

Make sure that the terminal screws are tight. Re-seal the cable gland by screwing the cover
back on.

9.

In order to avoid connection errors always take note of the hints given in the section
connection check!

4.1

Quick wiring guide

Terminal layout

Fig. 5:

Wiring the field transmitter

NOTICE

ESD - Electrostatic discharge
► Protect the terminals from electrostatic discharge. Failure to observe this may result in

destruction of parts of the electronics.

Advertising