-8-

6.1.2

What to do when the measurement head has been replaced

Please use the following procedure.

(1) Connection

Connect the detectors to the front of the column unit and the included power cord to
the back.

(2) Power on

Turn on the power switch in the rear of the unit (press the side marked with a circle)
to supply power (85-265V AC and 50/60Hz). The unit will start up in Measurement
Mode.

In Measurement Mode, hold down the left arrow (SET) key for two seconds, which
will switch to Set-up Mode.

(3) Inputting setting values

Please input the setting values. For details, please refer to 6.3 “Set-up mode.”

In Measurement Mode, hold down the right arrow (MAS) key for two seconds. After
that, pressing the key down once switches to Adjust Detector (ADJ).

(4) Adjust Detector

Please adjust the detector. For details, refer to 6.4.2 “Adjusting the detector.”

In “Adjust Detector” mode, holding down the ENT key for two seconds switches to
Master Matching.

(5) Master matching

Please match to a master. For details, refer to 6.4.1 “Master matching.”

(6) Measuring

In Measurement Mode, pushing the ENT key once switches to Hold Measured
Values. This can be cleared by pressing the RST key. However, if the master match
was unsuccessful, the measured values cannot be saved. For details, please refer to
6.2 “Measuring.”