-16-

6)

Small master matching will be performed, and
“MAX M.” will be displayed.

7)

Set the large master to a detector, and when the bar
display stabilizes, press the ENT key.

8)

Large master matching will be performed.

If there are no errors....... To 9)
If there is an error. .......... To 10)

9)

“MAS OK” will be shown in the eight-digit display
for two seconds, after which Master Matching will
be concluded and the unit will return to
Measurement Mode.

10) The following error messages may appear in the

eight-digit display.

“ERR ZERO”. . . A “zero” error (outside the

master matching range)

“ERR MAG” . . . A “sensitivity” error (outside the

master matching range)

“ERR REV” . . . . A MAX/MIN reversal
“ERR OFFR” . . . Out of measurable range
Press the ENT key to return to B above.
If the detector requires adjustment, press the right
arrow key and proceed to adjust the detector.

11) If master matching is being performed one at a

time, and master matching is OK for both items, the
unit will switch to a new master matching channel
and return to B above.

12) If the RST key is pressed during master matching,

the unit will return to its previous state. If “MIN
M.” was displayed, it will return to Measurement
Mode.

RST

ENT

MAS

SET

1

2

3

SET

MAS

6)

7)