6.3

Set-up mode

Set-up must be performed for each program used. After changing the program number,
please perform set-up.

6.3.1

Selecting the measurement range

1)

In Set-up Mode, hold down the left arrow key for at least two seconds.

2)

The orange “SET” LED will light up, “RANGE” will be displayed and the
system will be in Set-up Mode.

3)

Push the ENT key once.

4)

The current measurement range will be displayed.

5)

Press the up and down arrow keys to select a measurement range (5, 10, 20, 50,
100, 200, 500, and 1000 microns).

6)

After making a selection, press the ENT key.

7)

After measurement range selection has been completed, next is the set-up of the
shift value. “SHIFT” will now be displayed.

6.3.2

Setting the shift value

For one-sided tolerances and such, the bar display will move to the center. Set this
function to 0 to make it inactive.

1)

With “SHIFT” displayed, press the ENT key.

2)

The current shift value will be diaplayed.

3)

Use the up and down arrow keys to enter the new shift value.

4)

After entering the shift value, press the ENT key.

5)

After the shift value is set, next will come selection for mask matching. “MAS
SET” will be displayed.

6.3.3

Selection for mask matching

1)

With “MAS SET” displayed, press the ENT key.

2)

The current mask matching setting value will be displayed.

3)

Use the up and down arrow keys to select either “ALL” or “EACH.”

4)

When carrying out mask matching:
“ALL” . . . . . . Mask matching is performed all at once.
“EACH” . . . . Mask matches are performed one by one.

5)

After selecting the mask match settings, press the ENT key.

6)

After mask matching set-up is complete, next will come polarity selection for
each measuring unit. “XDUCER” will be displayed.