Operation Modes

15

3

Operational Modes

In this chapter…

Operation Modes

page 15

Output Signals

page 17

System Set Up

page 18

3.1

Operation Modes

The SafEye 700S has four operational modes:

Normal Mode, page 15

Maintenance Call Mode (3mA Output), page 15

Fault Mode, page 16

Zero Calibration Mode (1mA Output), page 16

3.1.1 Normal Mode

This mode is used for gas detection. In normal mode, the following statuses
are possible:

Normal: Signal received from gas detection is at safe levels

Warning: Gases have been detected at warning levels

Alarm: Gases have been detected at alarm levels

Note: For the standard 0-20mA output, the Warning and Alarm levels are
not relevant. The user chooses these alarm levels at the controller. The
output Detector will be 4mA at zero reading and 20mA for full-scale reading.

Warning and Alarm states are relevant where the stepped 0-20mA output is
used and will energize the warning and alarm relays.

If the RS485 output is used, the Detector changes its status from ‘N’ to ‘W’
at Warning level and to ‘A’ at Alarm.

3.1.2 Maintenance Call Mode (3mA Output)

This mode indicates low signal or low signal ratio that may be caused by a
dirty window; misalignment; poor source or that one of the Detectors
parameters is at the ‘limit’ value.

The Detector continues to function and reads any gas present, but provides
a (3mA) pre-warning signal that a maintenance procedure is required.