3 default detector set up, Default detector set up, Table 9: detector default set up – Spectrex Xenon 700 SIL Open-Path Gas Detection System User Manual

Page 33: Table 10: source default set-up

Advertising
background image

TM 799200 Rev (C), January 2014

System Set Up

21

The Source heated optic must be defined with the order in two options:

Heated continuously

Start heating below 41°F (5°C). (default)

3.3.3 Default Detector Set up

The Detector has 8 functions that can be programmed according to the
customer requirement at factory or at customer facility using a software
Host or a handheld unit. The standard set up is as follows:

Table 9: Detector Default Set Up

Function

Setting

Gas Type

2

Full scale sensitivity

5 LEL.m

BG Zero Calibration

Enabled

Accessory Relay

Warning*

Alarm Latching

No

0-20mA

Continuous**

Beam block during alarm

No latch

Heat mode

Auto

Heat on

5

* Only refers to wiring option with Accessory relay output

** Only refer to the wiring options with 0-20mA

Table 10: Source Default Set-Up

Function

Setting

Heat mode

Auto

Heat on

5

The Source default can be changed only at the factory.

Advertising