8000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Page 42

DC Current (Electrometer Mode) (Option)


Press DCI to enter the DCI measurement mode – the status display will indicate the settings and
range information.

Selecting ranges 10nA , 100nA, 1uA, 10uA will activate the electrometer mode. Inputs are measured
from the IMeas BNC terminal on the right hand side of the multimeter. An LED indicates the BNC
terminal set is active.

Status Display Section

Description

X RANGE

eg. *10mA DC

Display / Selection of DC Current range –

Use menu dial or

RANGE UP RANGE DOWN

buttons to change range.

Selected range indicated by a *

MANUAL

Ranging mode : AUTO or MANUAL

FRONT

FRONT or REAR terminals in use (Rear to 1A Max.)

Additional Settings

Description

DIGITS

Set measurement resolution

FILTER TIME

Set FILTER TIME :

250ms to 64s

Dynamic Mode for Automatic filter setting

based on measurement stability

ANALOGUE FILTER

Set ANALOGUE FILTER mode :

ON / OFF

READING MIN / MAX

Display READING MIN / MAX measurement data

for the selected function / range

TRIGGER

TRIGGER mode :

AUTO / MANUAL

GUARD TO LOW

Set GUARD TO LOW mode :

FLOATING / SIGNAL LOW

MATHS, NULL, dB

Set NULL : Store zero offset value

Clear NULL : Clear zero offset value

Set OFFSET : Add a user entered value to the measurement

Set SCALE : Multiply measurement by a user entered value

Clear FORMULA : Clear the formula entered (SCALE & OFFSET)

See dB function for use of Set dB, dB SET ZERO & dBm REF 600 Ohms

INTERNAL

TEMPERATURE

Displays the internal temperature