8000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Page 46

4-Wire Resistance


Press

Ω 4WIRE to enter the Ω 4WIRE measurement mode – the status display will indicate the

settings and range information.

Status Display Section

Description

X RANGE

eg. *1

Ω DC

Display / Selection of 4-Wire Resistance range –

Use menu dial or

RANGE UP RANGE DOWN

buttons to change range.

Selected range indicated by a *

MANUAL

Ranging mode : AUTO or MANUAL

FRONT

FRONT or REAR terminals in use

Additional Settings

Description

DIGITS

Set measurement resolution

FILTER TIME

Set FILTER TIME :

250ms to 64s

Dynamic Mode for Automatic filter setting

based on measurement stability

ANALOGUE FILTER

Set ANALOGUE FILTER mode :

ON / OFF

INPUT TERMINALS

Set INPUT TERMINALS to use :

FRONT / REAR

RATIO

Set RATIO mode :
ON / VALUE / OFF

READING MIN / MAX

Display READING MIN / MAX measurement data

for the selected function / range

TRIGGER

AUTO / MANUAL

GUARD TO LOW

Set GUARD TO LOW mode :

FLOATING / SIGNAL LOW

MATHS, NULL, dB

Set NULL : Store zero offset value

Clear NULL : Clear zero offset value

Set OFFSET : Add a user entered value to the measurement

Set SCALE : Multiply measurement by a user entered value

Clear FORMULA : Clear the formula entered (SCALE & OFFSET)

See dB function for use of Set dB, dB SET ZERO & dBm REF 600 Ohms

OHMS LOW CURRENT

Reduced measurement current to reduce power dissipation / self

heating during measurement

OFF / ON

OHMS COMPENSATION

Compensates for thermal voltages which may be present

at the measurement terminal connection

OFF / ON

POWER / CURRENT

Displays the power dissipation / current flowing

for the specific range & measurement

INTERNAL

TEMPERATURE

Displays the internal temperature