8000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Page 64

Calibration

Command Format

Parameters

Enter Password

k<Password>

Password = 1 to 10
digits

Set Password

K<Password> Requires

calibration

mode to be active


If a command includes a value which cannot be set due to, for example, the value being higher than

the range maximum, the multimeter will reject the command and stay set as it is (the multimeter will

return a *9 code).

The multimeter will respond to the commands sent with the response codes as detailed at the

beginning of this section. These codes can be used to ensure that potentially hazardous conditions

are clearly indicated to the operator and to maintain control of these outputs. This allows the

multimeter to be returned to a safe state once the testing required has been completed.

This functionality is employed within the ProCal calibration software from Transmille to allow safe

operation of the multimeter.Commands that change the calibration memory require the correct

calibration password to be sent.

Command Format Parameters

Set positive Cal Factor

CP

Input signal limits 70-130%

Set negative Cal Factor

CN

Input signal limits 70-130%

Set zero Cal Factor
Set AC 20% Cal Factor

CZ

DC: Input signal limits –5 to 5%,
value=0
AC: Input signal limits 5 to 30%,
Value=20%

Calibrate to Value

CV <value>

value = value of input signal in
current range units
Positive value sets positive cal
factor
Negative value sets negative cal
factor
Value Limits: 50-150% of range
Input signal limits: 50-150%
(Except AC 1kV 10-150%)
(AC 30A 10-150% )

Calibrate Zero Factor to Value

CA <value>

value = value of input signal in
current range units
Value Limits: -5 to 50% of range
Input signal limits: -5 to 50%

Set electrometer voltage output
calibration factor
(adjusts voltage output)
Stored once. Calibrate at 300V

CALIBRATE:ELOUTPUT
<space> <value>

value = measured electrometer
voltage output
limits: [Electrometer Voltage]
±25V

Set electrometer correction
calibration factor
(adjusts resistance calculation)
Stored for each voltage setting

CALIBRATE:ELMEASURED
<space> <value>

value = measured electrometer
voltage output
limits: [Electrometer Voltage]
±25V

Read back electrometer
correction calibration factor

CALIBRATE:ELMEASURED
<space> ?