Calibration – Transmille 8000 Series User Manual

Page 66

Advertising
background image

8000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Page 65

Calibration

Calibration Overview

To adjust the 8000 Series multimeter the multimeter can either be connected to a computer via the

available interfaces. Calibration constants stored within the multimeter can then be adjusted using

the built-in menus via the 8000 Series front panel. To prevent unauthorised access to calibration

menus, a password is required before access is granted. Adjustment can be completed without

disassembly of the multimeter.

WARNING

The information in this section is intended only for qualified personnel.
The user must at all times be adequately protected from electric shock.

Each function e.g. DC voltage, AC Current, Resistance etc. has several ranges. Each range has one

or more calibration constants. See table below.

The 8000 Series allows any calibration constant to be adjusted independently of any other, therefore

it is possible to adjust a single range without needing to adjust any other points. Altering the

calibration constants directly changes the multimeter measurement. Adjusting the multimeter simply

involves changing the constant until the output reads correctly.

Entering Calibration Mode


Press MENU

Press

Æ


Enter password (default = Model Number, ie. 8081 or 8071)

Press ENTER

Multimeter will display CAL PASSWORD OK if successful

Advertising