8000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Page 56

Shunt Measurement : Multimeter Setup (Option)


The precision shunt measurement function allows accurate measurement of current shunts by
monitoring the current supplied to the shunt and voltage measured across the shunt. By monitoring
both the current and voltage the 8000 Series can provide an accurate measurement of the current
shunt connected.

The diagram shows the connection configuration required, using a 3000 Series to provide the 30A
current source and both the current and voltage measured by the 8000 Series. When SHUNT
MEASUREMENT mode is selected, the 8000 Series will automatically switch between voltage and
current measurement terminals in a continuous cycle until shunt mode is exited.

Press USER then press

Ä until SHUNT MEASUREMENT is displayed


Press ENTER to select SHUNT MEASUREMENT

The secondary display can be used to display the separate Voltage and Current readings by using
the menu dial to scroll to this display.

Additional Settings

Description

DIGITS

Set measurement resolution

FILTER TIME

Set FILTER TIME :

250ms to 64s

Dynamic Mode for Automatic filter setting

based on measurement stability

ANALOGUE FILTER

Set ANALOGUE FILTER mode :

ON / OFF

INPUT TERMINALS

Set INPUT TERMINALS to use :

FRONT / REAR

READING MIN / MAX

Display READING MIN / MAX measurement data

for the selected function / range

TRIGGER

AUTO / MANUAL

VOLTAGE / CURRENT

DISPLAY

Displays the measured VOLTAGE and CURRENT separately as

well as the calculated WATTS on the main display

GUARD TO LOW

Set GUARD TO LOW mode :

FLOATING / SIGNAL LOW

INTERNAL

TEMPERATURE

Displays the internal temperature