Transmille 8000 Series User Manual

Page 58

Advertising
background image

8000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Page 57

DC Power / Dual Input V/30A (Option)


The DC Power / Dual Input V/30A function allows the 8000 Series to automatically switch between
the voltage and current inputs and display a calculated value in Watts.

Press USER then press

Ä until DC POWER DUAL INPUT V/30A is displayed


Press ENTER to select DC POWER DUAL INPUT V/30A

The 8000 Series will automatically switch between the Voltage and High Current Terminals,
separately measuring both inputs and displaying a calculated value in Watts on the main display.

The secondary display can be used to display the separate Voltage and Current readings by using
the menu dial to scroll to this display.

Additional Settings

Description

DIGITS

Set measurement resolution

FILTER TIME

Set FILTER TIME :

250ms to 64s

Dynamic Mode for Automatic filter setting

based on measurement stability

ANALOGUE FILTER

Set ANALOGUE FILTER mode :

ON / OFF

INPUT TERMINALS

Set INPUT TERMINALS to use :

FRONT / REAR

READING MIN / MAX

Display READING MIN / MAX measurement data

for the selected function / range

TRIGGER

AUTO / MANUAL

GUARD TO LOW

Set GUARD TO LOW mode :

FLOATING / SIGNAL LOW

VOLTAGE / CURRENT

DISPLAY

Displays the measured Voltage and Current separately as well as

the calculated Watts on the main display

INTERNAL

TEMPERATURE

Displays the internal temperature

Advertising