Safety Information

-14-

CAUTION

To prevent the risk of infection, wipe the forehead rest with disinfectant
ethanol for each patient. For details on how to disinfect, consult a specialist.
The forehead rest may be corroded if a disinfectant other than those above is
used.

CAUTION

To ensure cleanliness, replace the chin rest paper for each patient.

CAUTION

When adjusting the forward position of the main unit, be sure to move the
main unit slowly toward the patient while looking from the side of the patient,
to prevent accidental contact of the objective lens with the patient.

CAUTION

Do not place your hands or fingers or allow the patient to place his/her hands
or fingers between the stage and base. Otherwise, injury may occur.

CAUTION

When the instrument is not going to be used, turn OFF the power.
Also, unplug the power cable from the AC outlet when it is not going to be
used for long periods of time.

CAUTION

The instrument weighs approximately 26 kg (57 Ibs). Be sure that at least two
people lift the instrument to transport it and that it is lifted by gripping the
indentations for lifting. Otherwise, injury may occur.

When Problem Occurs

WARNING

Should any of the following occur, immediately turn OFF the power of each
instrument, unplug the power cable from the AC outlet, and contact your sales
representative or local Canon dealer.
• When there is smoke, an odd smell or abnormal sound.
• When liquid has been spilled into the instrument or a metal object has

entered through an opening.

• When the instrument has been dropped and it is damaged.

Maintenance and Inspection

WARNING

For safety reasons, be sure to turn OFF the power when the inspections
indicated in this manual are going to be performed. Otherwise, electric shock
may result.

WARNING

When the instrument is going to be cleaned, be sure to turn OFF the power,
and unplug the power cable from the AC outlet.
Never use alcohol, benzine, thinner or any other flammable cleaning agents.
Otherwise, fire or electric shock may result.

WARNING

Clean the plug of the power cable periodically by unplugging it from the AC
outlet and removing dust or dirt from the plug, its periphery and AC outlet with
a dry cloth. If the cable is kept plugged in for a long time in a dusty, humid or
sooty place, dust around the plug will attract moisture, and this could cause
insulation failure which could result in a fire.