Electromagnetic Immunity

The CF-1 is intended for use in the electromagnetic environment specified below.
The user of the CF-1 should assure that it is used in such an environment.

Immunity Test

IEC 60601 Test

Level

Compliance

Level

Electromagnetic Environment –

Guidance

Electrostatic dis-
charge (ESD)

EN IEC 61000-4-2

±(2, 4, 6) kV
contact

±(2, 4, 8) kV air

±(2, 4, 6) kV
contact

±(2, 4, 8) kV air

Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.

Electrical fast tran-
sient/burst

EN IEC 61000-4-4

±2 kV for power
supply lines

±1 kV for input/
output lines

±2 kV for power
supply lines

±1 kV for input/
output lines

Mains power quality should be that of a
typical commercial or hospital environ-
ment.

Surge

EN IEC 61000-4-5

±1 kV differen-
tial mode

±2 kV common
mode

±1 kV differen-
tial mode

±2 kV common
mode

Mains power quality should be that of a
typical commercial or hospital environ-
ment.

Voltage dips, short
interruptions and
voltage variations
on power supply
input lines

EN IEC 61000-4-11

<5% U

T

(>95% dip in
U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in
U

T

) for 5 sec

<5% U

T

(>95% dip in
U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in
U

T

) for 5 sec

Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the CF-1 requires
continued operation during power
mains interruptions, it is recommended
that the CF-1 be powered from an
uninterruptible power supply or a bat-
tery.

Power frequency
(50/60 Hz)
magnetic field

EN IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.

NOTE: U

T

is the AC mains voltage prior to application of the test level.