MS-70 / MX-50 / MF-50 / ML-50

23

1. Turn on the analyzer.

The gram unit (of weighing mode) is displayed.


Enter the Quick Mode

2. Press

the

SELECT

key to display an analysis

mode and press the

or

key to select

quc

.




Select ACCURACY

3. Press

the

SELECT

key to select

ACCURACY.

When

ACCURACY

is selected, HI, MID. or LO blinks.

4. Press

the

or

key to select LO of ACCURACY.




Set Drying Temperature at the Sample Pan

5. Press

the

SELECT

key to select the drying

temperature.

6. Press

the

or

key to set 130 °C.



Store Parameters and Finish the Operation

7. Press

the

ENTER

key to store the parameters.

The weighing mode is automatically displayed.






Put a Sample on the Pan

8. Put the breeze break ring, pan support, pan handle

and sample pan in order. (With no sample.)


9. Close the heater cover.

10. When displaying a stable value, Press the

RESET

key.

Avoid mechanical vibration, breeze and environmental
noise during measurement. If it deviates from zero
display, press the

RESET

key.

Close

Preparation

Quick mode

ACCURACY

Temperature

Quick drying

ML-50 does not display