Step drying – A&D ML-50 User Manual

Page 39

Advertising
background image

MS-70 / MX-50 / MF-50 / ML-50

37

7.2.3. Step

Drying

Standard Mode, Automatic Mode or Manual Mode

Refer to page 39 for Timer Mode


This explanation uses the following parameters and sample displays of MX-50.






"Temperature 1" can be set higher than "temperature 2" in step drying.

Common

Items

Program

number ..........................................5 ( PROG 5 )

Drying

program

Heating pattern....................Step drying (

)

Drying temperature 1 ..........120°C

Drying temperature 2 ..........160°C

Time.....................................5.0 minutes

Measurement

Unit..............................................Moisture content

Minimum scale value during measurement ..0.01 %

Minimum scale value of gram display...........0.001 g

Sample

quantity............................................Approximately 5 g

Data

memory

function ..................................Not used

Items for Standard Mode

Analysis

mode ..............................................Standard mode (Symbol: 5td )

ACCURACY .................................................MID.

Analyzing mode to finish measurement............Automatic setting by ACCURACY

Items

for

Automatic

Mode

Analysis

mode ..............................................Automatic mode (Symbol: U-a )

Analyzing mode to finish measurement............0.05 %/min

Items for Manual Mode

Analysis mode ..............................................Manual mode (Symbol: U-m )



1. Display gram unit (of the weighing mode).

Select a Program Number to Edit the Measurement Program

2. Press

the

PROGRAM

key and press the

or

key to select a program number.

3. Press

the

ENTER

key to use the number.

4. The analyzer displays

end

and returns to the

weighing mode.

Caution If the data memory function is active, the data

memory number (MEM) is displayed in place of the
program number (PROG).

% MOIST

D

/W

/D

PROG 1

0.000

g

PROG "5

end

PROG 5

0.000

g

Program number

Time

Drying temperature 2

Drying temperature 1

Stage 1

Stage 2

(ML-50 can not select step drying)

Advertising
This manual is related to the following products: