Chapter 5 _____________________________________________________________ User Interface

4. Select

Quantity and press

Set to select the quantity.

5.

Select the quantity and press

Select. Only one active quantity

can be selected at a time. To change which quantities are active,
go to Menu - Display - Quantities and units.

6.

Select the first limit and press

Set (if the alarm function is

on, turn it off). Use the arrow buttons to set the alarm level. To

select the sign for the alarm level, press the

+/- button. To

save the settings, press

OK.

7.

Select the second limit and follow the instructions in step 6. The
alarm will go off when either the upper or lower limit is
exceeded.

If you will use only one limit, for example, if you want to detect
if the dewpoint will rise above -40

°

C, do the following:


Set the upper limit to -40

°

C and the value of the lower limit so

low that it will never be reached, for example, to -80

°

C. Thus,

DM70 will alert when T

d

rises above

-40

°

C.

8.

To activate the alarm and to return to the basic display, select

Alarm On/Off and press

On. The note symbol

♫ will appear

in the upper left corner of the display.

9.

When the alarm level is reached, you can stop the alarm by

pressing

OK. To reactivate the alarm, press

Yes. To

completely stop the alarm, press

No.

NOTE

The alarm does not work when the device is turned off.

VAISALA _______________________________________________________________________ 25