Temperature, Ppm (ppmv or ppmw), Ppm (ppm – Vaisala DM70 User Manual

Page 85: Or ppm

Advertising
background image

Chapter 11 ____________________________________________________ Technical Specifications

DMP74A
Response time: 63 % (90 %) at flow rate 0.2 m/s, 1 bar
absolute pressure, +20 ˚C
0

Æ-40 ˚C T

d

(32 Æ40 ˚F T

d

)

20 s

(120 s)

-40

Æ0 ˚C T

d

(-40 Æ32 ˚F T

d

) 10 s

(20 s)

DMP74B/C
Response time: 63 % (90 %) at flow rate 0.2 m/s, 1 bar
absolute pressure, +20 ˚C
0

Æ -60 ˚C T

d

(32 Æ76 ˚F T

d

) 50 s

(340 s)

-60

Æ0 ˚C T

d

(-76 Æ32 ˚F T

d

) 10 s

(20 s)

Temperature

Measurement range

-10 ... 60 ˚C (14 ... 140˚F)

Accuracy at +20 ˚C ±0.2

˚C (±0.36˚F)

Typical temperature
dependence of electronics


±0.005 ˚C/˚C

Temperature sensor

Pt100 IEC751 1/3 class B

PPM (ppm

v

or ppm

w

)

Measurement range (typical)

DMP74A

40 ... 200000 ppm

DMP74B/C

10 ... 20000 ppm

Accuracy at +20 ˚C (+68 ˚F)

< 40 ppm

± (0.5 ppm + 25.4 % of
reading)

> 40 ppm

± (7.3 ppm + 8.3 % of
reading)

VAISALA _______________________________________________________________________ 83