To same as ti/ii, Point adjustment, To same as t – Vaisala DM70 User Manual

Page 78: I/ii

Advertising
background image

USER'S GUIDE____________________________________________________________________

4.

From the list of parameters, select T

I/II

5.

At this point you need to select the adjustment method. If the
reference probe is connected, two options are displayed: To
same as T

I/II

, 1-point adjustment or 2-point adjustment.

6.

You are prompted to check the settings, make sure that the
pressure units of DM70 and DMT340 are the same. If not,

change the unit by pressing

Unit on the corresponding row.

To Same as T

I/II

NOTE

This adjustment method automatically uses the reading of the
reference probe as the reference value for making the adjustment.

7.

After selecting To same as T

I/II

, confirm the adjustment by

pressing

Yes.

If you press

No, a note is displayed on the screen reminding

you to follow the adjustment procedure described in the User's

Guide. Press

OK and the adjustment mode display returns

and no changes are made.

1-Point Adjustment

7.

After selecting 1-point adjustment, press

Select to

continue.

8.

Set the probe to the reference temperature. You can follow the

stabilization from the graph display by pressing

Graph.

When the reading has stabilized, press

Ready.

9.

Use the arrow buttons to give the reference temperature value

and then press

OK.

10.

Confirm the adjustment by pressing

Yes. (If you press

No, the adjustment mode display returns and no changes are
made.) The adjustment has now been completed.

11.

To exit the adjustment mode, press

Back.

12.

To return to the basic display, press

Exit.