USER'S GUIDE____________________________________________________________________

12.

Confirm the adjustment by pressing

Yes. If you press

No,

the adjustment display returns and no changes are made.

13.

Now the adjustment is completed. To exit the adjustment mode,

press

Back.

14.

To return to the basic display, press

Exit.


1) If there is a temperature difference between the generated reference

and the probe temperature, relative humidity must be calculated to
correspond to the temperature of the probe.

Adjusting Dewpoint T

d/f

NOTE

For DMP74B/C, the sensor purge must be performed manually one
hour before adjusting the dewpoint.

T

d/f

Adjustment of DM70

For DMP74A, the reference temperature of the dewpoint should be -
40 ˚C ... -60 ˚C (-40 ˚F ... -76 ˚F).

For DMP74B/C, the reference temperature of the dewpoint should be
-57 ˚C ... -67 ˚C (-70.6 ˚F ... -88.6 ˚F).

To adjust the dewpoint, follow the instructions below:

1.

Set the probe to the reference condition. Follow the stabilization
from the graphic display by pressing

Graph. The minimum recommended stabilization time for

this critical adjustment is 5 hours.

2.

Perform the manual purge if necessary.

3. Press

the

Calibration button of the probe handle (see number 4

in Figure 1 on page 11) using a tool with a thin, sharp point, for
example, a small screwdriver. The indicator will switch into the

adjustment mode. To confirm calibration, press

OK.

4. Select

T

d/f

and press

Select.

5.

To check the environmental settings, press

Yes. To go to

calibration directly, press

No.