Adjustment of dmt340 series transmitters using, Dm70 – Vaisala DM70 User Manual

Page 73

Advertising
background image

Chapter 9 _______________________________________ Calibration, Adjustment, and Maintenance

6. To

select

1-point adjustment, press

Adjust and then

Select.

7.

When the reading has stabilized at the reference level, press

Ready.

8.

Use the arrow buttons to give the reference value. Then press

OK.

9.

Confirm the adjustment by pressing

Yes. If you press

No,

the adjustment mode display returns and no changes are made.

10. Wait until the automatic auto-calibration has taken place.

11. When the calibration is completed, press

Back to exit the

adjustment mode.

12. To return to the basic display, press

Exit.

13. Place the screw back onto the calibration button.

14. To seal the calibration, attach a sticker on the calibration button.

NOTE

Several AutoCals can be necessary after this adjustment has been
performed until the transmitter reaches full accuracy.

T

d/f

Adjustment of DMT340 Series Transmitters

using DM70

NOTE

For DMT340 series transmitters with the M-sensor, the sensor purge
must be performed manually one hour before adjusting the dewpoint.

For DMT340 series transmitters, the reference temperature of the
dewpoint should be -57 ˚C ... -67 ˚C (-70.6 ˚F ... -88.6 ˚F).

To make dewpoint adjustment of DMT340, follow the eight first steps
below and then continue according to the chosen adjustment method:

1.

Turn off DM70.

2.

Connect the cable 211339 between DMT340 (service port) and
DM70 (connector I or II). Turn on both the devices.

3.

Set the probe to the reference condition. Follow the stabilization
from the graphic display by pressing

VAISALA _______________________________________________________________________ 71